by Mark Heyda,
MBH Engineering Systems
Hanna
pH and ORP Meters
This monograph is
your guide to using a new pH meter. If you have purchased
an ORP electrode, this article will address that as well.
Your new pH meter may be very simple with automatic
features, or a bit more complex with many advanced
features. Either way, the same basic procedures need to
be followed. These instructions are tailored to Hanna
Instrument pH meters but can be applied to most meters.
Understanding pH and
ORP electrodes and calibration
The electrode
itself or sensor sends a signal to the digital meter based
on the acidity, alkalinity or ionic potential (ORP) of the
solution. The signal is a very weak voltage measured in
millivolts. The meter converts the signal to pH and
displays the result. However the signal can and will change
over time for all electrodes. This is a good analogy: Think
of the electrode as a chemical battery. Over time the
battery no longer develops the same voltage and the attached
flashlight bulb light weakens. Eventually the light fails
totally. The same is true for an electrode. Because the
signal for the same pH value changes, the meter needs some
way to assign the proper pH value to it. This is the reason
why pH meters need periodic calibration. Each time the
meter is calibrated, the meter assigns a new set of values
to the incoming signals. For instance, a new pH electrode
when placed in a known standard solution (commonly referred
to as a “buffer”) with a value of pH 7.01 may have an output
signal of -5.1mV. The meter after calibration will show on
its display a reading of 7.01 (+/- the accuracy of the
meter’s electronics). Let us assume that after a month the
same electrode now sends a signal of -5.5mV to the meter for
the same buffer. Without recalibration, the meter will show
the pH incorrectly. The error of 0.4mV is the drift of the
signal. In this case the error is small so the offset from
the true value will not be very much (0.01pH). However, as
the electrode ages, the error will increase and the rate of
the change will increase as well. That is why electrodes
need more frequent calibration as they age. At the end of
their useful life, the electrode may exhibit an offset of
-30mV which is equivalent to 0.5 pH units or more depending
upon where on the pH scale the reading is occurring. To the
user, all of this happens automatically each time the meter
is calibrated. (No math required!) However, when the drift
exceeds about -30 to -35mV, the electrode becomes unstable.
The user will find that the pH value on the display changes
for no reason with the same solution. And, the meter more
than likely will not calibrate at all.
Two or More Point
Calibration
Now that we
understand the basics of calibration, why should a meter be
calibrated at more than one pH value? Most pH meters should
be calibrated at a minimum of two points. The most common
points are values of pH 7.01 and 4.01 or 7.01 and 10.01 if
you are going to be reading alkaline values. The more
advanced Hanna meters will let the user calibrate at three,
four or five pH values. These meters will also let the user
select which buffers are to be used. While the majority of
meters have memorized values of 4.01, 7.01, 10.01 as well as
European buffer values (1.68, 6.86, 9.18), custom buffers
are available in pH values from 1.00 to 13.00. To achieve
the best accuracy, the calibration should cover the range of
the desired measurement values. For instance if the user
wanted to measure a solution which had values expected to
range between 3.00 and 4.00, one would want to calibrate at
3.00 and 7.01 rather than 4.01 and 7.01. This eliminates
the uncertainty associated with trying to extrapolate a
value beyond the calibrated values. Some very accurate
Hanna meters (both bench top and portable) will be able to
read and calibrate (using special buffers) to a thousandth
of a pH, i.e. 3.204 +/- 0.002 pH. Getting back to the
science behind a multi-point calibration requires an
understanding of “slope”. Earlier we talked about offset.
If the pH values were linear, i.e. a pH value could be
calculated by the meter using 59.16mV per pH unit. The
meter would display the proper value once the offset error
was adjusted by the calibration. However, to make life more
complex, the slope of the pH line when plotted (pH vs. mV)
may be altered at pH values which vary from pH 7. Hence the
error in the reading is larger as the reading increases or
decreases from neutrality (pH 7). The good news is that the
multi-point calibration also corrects for this deviation.
The resulting readings then are correct at all values
between the calibration points. If the readings go beyond
the calibrated range, the meter assumes linearity and simply
extrapolates the value to be displayed. The true value may
be slightly different. Much depends upon the users need for
accuracy at the measured value and the need for a more
advanced meter.
Temperature
Compensation
One more point
needs to be considered. The temperature of a solution when
depressed or elevated from 20°C will also contribute an
error to the pH reading if the meter does not have a
temperature sensor built in to correct for this variation.
The larger the divergence becomes, the greater the error
will be. The error only becomes significant if the pH value
diverges from neutrality but increases as the pH values
become more alkaline or acid. In other words at pH 9, the
error increases with the same temperature offset versus pH
8. Most pH meters today include a temperature compensation
function and will automatically correct for the error caused
by temperature variations.
Technique for
Calibration
If the user
follows these simple steps, the meter will be calibrated
properly.
1.
Rinse the pH electrode with tap water.
2.
Dry the outside of the electrode but blot the bottom
of the sensitive glass with a paper towel. This is done to
avoid abrasion of the glass measuring end and to prevent
dilution of subsequent use.
3.
Place the electrode into a pH 7.01 buffer solution
making sure that the junction (located on the bottom or side
of the electrode) is wet. For most electrodes this means an
immersion of only 1 to 2 inches.
4.
For more precise lab work, use a magnetic stir plate.
5.
Turn the meter on and select the calibration mode.
6.
The meter will normally default to asking for a pH
7.01 buffer. If not use the up and down keys to select pH
7.
7.
The meter’s display will ask to “CONFIRM or CFM” when
the value in the display has stopped changing. If the
electrode is in good condition, the value will jump to near
the buffer value very rapidly. As the electrode
deteriorates or if dirty/clogged, this process will take
longer.
8.
Press the proper key to CFM or other key as stated in
your manual.
9.
The meter will now ask the user to place in another
buffer (usually pH 4.01). The user can accept this value or
change it to the desired second calibration point.
10.
Rinse the electrode as before in tap water and dry as
in step 2.
11.
Place the electrode in the buffer selected.
12.
Wait for the meter to accept the second value and
press the confirm key when prompted.
13.
Repeat as needed for additional buffers if needed.
14.
If the meter indicates “Wrong buffer” it means that
the value selected does not match the immersion solution or
the electrode’s diverges too far from an accepted value. To
remedy, a new electrode must be used or the electrode must
be cleaned prior to use.
15.
If the meter indicates “Wong electrode” it could mean
that the attached electrode is of the wrong type (i.e.- ORP
rather than pH).
16.
Once the calibration at two or more points has been
completed, the meter will automatically return to the normal
measuring mode. It will show the temperature compensated
value of the solution. Some small variation may occur
depending upon the inherent accuracy of the meter and the
variation of temperature. If the probe and the solution are
at different temperatures, readings will continue to change
slightly as the two reach equilibrium.
Electrode Fouling
If an electrode
is dirty or clogged with oil, grease, debris etc., the meter
will still yield a digital result. The reading more than
likely will not be correct. Older electrodes which have
been extensively used in solutions other than plain water or
at elevated temperatures can be damaged. Sometimes even an
inspection of the electrode tip will not reveal any outward
problem. Cleaning solutions developed by Hanna Instruments
can be used to rectify these problems. Many solutions are
available for different industries and maladies. It is best
to consult with a knowledgeable distributor such as MBH
Engineering Systems to select the best solution(s) for your
application. The use of these products can easily extend
the useful life of an electrode.
Smart meters
developed by Hanna can identify electrode age and fouling.
These instruments are referred to as “Calibration Check™”
meters. The CAL-CHECK feature shows the user the condition
of the attached electrode after each calibration. This
allows users to know in advance whether a displayed value is
correct or marginal. It also should be a prompt to clean
the electrode. In fact some of the meters actually flash
“clean electrode” in the display!
Electrode Care
To prolong the
life of the electrode, it should be immersed in a special
solution when not in use. The solution is simply called,
Storage Solution (HI 70300). The benefits are twofold.
First, the glass will remain hydrated. This is very
important as a dry electrode will degrade and must be
rehydrated prior to use. Second, the salts in the solution
are at the same concentration as the salts inside the
electrode. Therefore no transportation of salts will occur
to deplete the pH reference. Technically one could say that
there is an absence of osmotic pressure across the
junction. The life of an electrode will vary depending upon
the type of solutions, temperatures, immersion times and
cleaning cycles. Typically electrodes will last about a
year although some can remain functional for two years.
Rarely do electrodes exceed this life span. In some cases
with severe conditions of use, electrode life may be as
short as one month or even a few days! Selection of the
proper electrode is important as hundreds of different types
of probes are manufactured. Each probe is designed to
accommodate different conditions of use to maximize its
useful life.
ORP Electrodes
ORP or
oxidation-reduction potential electrodes look similar to a
pH electrode but in fact function quite differently and are
not interchangeable with a pH electrode. Typically they
have a platinum tip although sometimes a gold tip can be
used. The voltage developed by these electrodes does not
drift nearly to the extent that their pH probe cousins do.
In other words these probes have a longer life. While ORP
probes are not calibrated as pH probes are, they should be
checked to see if their error is unacceptable. As a rule of
thumb, a divergence of more than 50mV from a standard
solution (HI 7022L 470mV) would show that the electrode
needs replacement. It is a good idea to condition an ORP
electrode prior to use. Hanna Instruments has two solutions
which can be used to help the tip respond properly to ORP
changes. These pre-treatment solutions are: HI 7091 or HI
7092 (reducing or oxidizing respectively). ORP electrodes
also respond well to cleaning solutions and need to be
immersed in “Storage Solution” when not in use.
Troubleshooting
What happens if
my meter is not working properly? Chances are that the
electrode is no longer serviceable. In the absence of a
Calibration Check™ meter, a quick easy method to determine
electrode failure is to look at the electrode offset. Some
meters and controllers have either a CAL Data key or a GLP
(good laboratory practice) key. Using this key, the meter
will show the last calibration time and date as well as the
offset, slope and buffers used with the calibration. If the
offset is numerically greater than [-30mV] i.e.: [-36mV],
the probe should be replaced. If your meter does not have
that feature, but has a RANGE key, you can determine the
offset as well. Simple place the electrode in a pH 7 buffer
and read the raw reading from the probe in millivolts by
switching the reading from pH to mV. The meter will show
the offset. In rare cases, the offset is acceptable but
meter is still not functioning properly, the slope may be
too large. However, in that case the meter more than likely
will not calibrate.
Final Thoughts
Hanna Instruments
continues to innovate and build what many refer to as the
"Cadillac" of pH meters and electrodes. The choices
available now are extensive for those in the market for a
meter. From the very excellent rechargeable portable field
instruments which are waterproof-- of course-- to excellent
research grade bench meters, Hanna manufactures all of
them. Because they build everything in their own factories,
they continue to offer great prices as well. Advanced
features like: intelligent electrodes, ISE capabilities,
Calibration Check™, millesimal pH ranges, GLP, data logging,
graphical displays, inductive recharging, auto-end points
and more are standard features on selected models. I have
seen meters in the field which are still in service after
twenty years. It is nice to know you are going to be
getting a reliable instrument with an iron clad no nonsense
warranty.
The above article may not be copied or
reproduced without consent of the author. All rights reserved. Mark Heyda
2008