Portable pH Meters
Waterproof Battery Powered pH/ORP & ISE Meters

 
With more than 50 different models to choose from, we have a
pH meter to fit
your application. All the pH meters we supply include a pH electrode with
integral or separate temperature compensation probes:
Features:
- Waterproof cases
- Plastic, stainless steel or titanium
electrodes
- Temperature compensation
- Automatic buffer recognition
- Rechargeable inductive batteries (select models)
- Data Logging (select models only)
- Built in printer (select models only)
- Field travel cases included
- Calibration check
- Calibration stability indicator and auto-end feature
- Sensor check or Cal-Check
- Hold/Data point capture
- Simple two button design or more flexible
multi-function designs
- mV scales for ORP measurement and pH sensor
diagnostics (select models only)
- Combination pH, ORP, temperature probe (select
models only)
- AC power via 12Vdc power supply (select models only)
- Back lighted display for low light conditions
(select models only)
- Computer output to download logged data (select
models only)
- "Smart" pH electrodes memorizes calibration data
With all these choices, we understand that choosing the
pH meter which will serve your needs best may present a challenge. The pH
meters are listed below in detail and we can certainly help if you want us to
personally assist you. However, we have provided a comparison table to
show the differences. All the pH meters have resolution of 0.01 pH (some
0.001 pH) and 0.1ºC
and are very easy to use (even if they have a lot of buttons).
These meters have menu based automatic easy calibration. Some meters have
other features which may be important and are unique to that product.
Note: pH tester HI 98127 only has resolution to 0.1 pH and the accuracy of the
pH tester HI 98128 is only +/- 0.05.
| |
Cal-check |
mV |
ORP/pH probe |
GLP |
ISE
|
"Smart" Probe |
Built-in Temp Sensor |
Metal Probe |
Backlit Display |
Logging |
Printing and Logging |
PC Output |
12V Power Supply Option |
| 9026 |
√ |
√ |
|
|
|
|
|
|
√ |
|
|
|
|
| 991002 |
|
√ |
|
|
|
|
√ |
√ |
|
|
|
|
|
| 991003 |
|
√ |
√ |
|
|
|
√ |
√ |
|
|
|
|
|
| 98140 |
|
|
|
√ |
|
√ |
√ |
|
√ |
√ |
|
√ |
√ |
| 98150 |
|
√ |
|
√ |
|
√ |
√ |
|
√ |
√ |
|
√ |
√ |
| 99161 |
|
|
|
|
|
|
√ |
|
|
|
|
|
|
| 99131 |
|
|
|
|
|
|
√ |
√ |
|
|
|
|
|
| 99141 |
|
|
|
|
|
|
√ |
√ |
|
|
|
|
|
| 98230 |
|
|
|
√ |
|
√ |
√ |
|
√ |
√ |
√ |
√ |
√ |
| 98240 |
|
√ |
|
√ |
|
√ |
√ |
|
√ |
√ |
√ |
√ |
√ |
| 9025 |
|
√ |
|
|
|
|
|
|
|
|
|
|
|
| 98127 |
|
|
|
|
|
|
√ |
|
|
|
|
|
|
| 98128 |
|
|
|
|
|
|
√ |
|
|
|
|
|
|
| 98170 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 98171 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 98172 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| 98180 |
√ |
|
|
√ |
|
|
|
|
√ |
√ |
|
√ |
|
| 98181 |
√ |
√ |
|
√ |
|
|
|
|
√ |
√ |
|
√ |
|
| 98182 |
√ |
√ |
|
√ |
|
|
|
|
√ |
√ |
|
√ |
|
| 98183 |
√ |
√ |
|
√ |
|
|
√ |
|
√ |
√ |
|
√ |
|
| 98184 |
√ |
√ |
|
√ |
√ |
|
√ |
|
√ |
√ |
|
√ |
|
| 98185 |
√ |
√ |
|
√ |
√ |
|
√ |
|
√ |
√ |
|
√ |
|
| 99121 |
|
|
|
|
|
|
√ |
|
|
|
|
|
|
| 99171 |
|
√ |
|
|
|
|
√ |
|
|
|
|
|
|

We are very excited about the new Cal-Check
Portable HI 9126--- a great waterproof pH meter. We use this pH meter for
field verifications. The 9126 is the next generation of the field proven
model HI 9025 that has been the "gold standard" for so many years.
This pH meter has the ability to calibrate
automatically using the standard pH buffers and one custom (user defined)
buffers to take advantage our any of the "technical" special buffers which cover
the entire pH range. (see accessories in this pH section). It also
has an option to use a back lighted display for low light environments.
The Cal-check feature shows the condition of the
electrode using a bar graph in the lower left of the display. Read more
about Cal-check in the "technical info" section of this website.
This meter includes a hard plastic travel case, pH
electrode, temperature sensor, AA batteries, buffers and instruction manual.
Click on picture for more information and specifications.
$395
Field Monitoring Meters for pH, ORP and
ISE: HI 9817x
And now, the 9026 has some big brothers: Introducing
the field monitoring Calibration Check™
Meters:
HI 98170 / HI 98171 / HI 98172. These pH
meters feature rechargeable batteries with an inductive recharger. The
battery life is about 200 hours for long operation in the field. But when
the batteries are low, simply put the meter in Hanna's inductive recharger
cradle for a few hours and you are ready to go. And to make life even
better, the recharger can be powered from a car 12 volt outlet or with the
included 115V adaptor.

While the rechargeable feature is nice, the real story
about this trio of meters are the powerful analytical assets: GLP, pH/ORP/ISE
measurements, ATC, auto-calibration at up to five points including custom buffer
values (user configured) plus seven standard values, backlit display, log
on demand, RS-232 interface and of course the Hanna calibration check 5 bar
graphic indicator which determines the electrode condition. Depending upon
which model is selected, the user can use a variety of electrodes such as: pH,
ORP, Fluoride, Sodium, and other ISE's.
Click on the picture on the right for detailed
specifications.>>>
Click to buy on-line:

The Next Evolution in pH meters:
Graphic
LCD Portable Meters: HI 98180, HI 98181, HI 98182, HI 98183, HI 98184 & HI 98185
These meters feature a rugged rubberized waterproof
"military" case, USB computer connectivity, backlit graphical display,
Calibration Check™ messages, logging of data, rechargeable batteries with an
inductive charger, accuracy 0.01 or 0.002 pH and 0.2mV. The last two
in the series will even accept ISE electrodes and read directly in ppm (mg/l).
These meters represent the next step in pH meter evolution. Who needs a
manual? The meters feature an extensive help menu covering every aspect of the owner's manual. In this
family of meters, Hanna Instruments also introduces advanced
conductivity/resistivity and dissolved oxygen products. HI 98180 pH only,
HI 98181 pH and mV 0.01pH/0.1mV, HI 98182 pH and mV- 200 data points- 0.001pH,
0.1mV and the HI 98183 has the features of the 98182 but uses a combined pH &
temperature sensor electrode.
ISE Meters: The HI 98184 and HI 98185
improve upon the HI 98183 by including ISE electrode capability. Using any
of the family of Ion Specific Electrodes, these meters will read directly in ppm
or mg/l with either monovalent or divalent ions. Logging points are
increased to 300 on demand data samples. Data can be downloaded thru the
waterproof USB port. Hanna software makes it easy.
Click on the above picture for a full color brochure
with detailed specifications or:
HI 98180 Series.
Click to buy on-line: HI 98181

Click to buy on-line: HI 98182

Click to buy on-line: HI 98185

Please call us for other meters in this line.
Plating Bath pH Meters and
Boiler & Cooling Tower pH Meters The
new Boiler and Cooling Tower pH Meter HI 99141 and the companion Plating Bath
pH
Meter HI 99131 feature two button control with Titanium body electrodes and
built-in Automatic Temperature Compensation. Click on picture for more specifications. The 99141
(Boiler/Cooling tower) has a flat tip and the 99131 (Plating) has a
recessed tip, otherwise they are the same. Both meters have accuracy of
0.01pH and the probes can be used in very hot liquids up to 221F which are
typical in these applications. Of course they are waterproof as well.
Calibration is a snap: both meters have a push button automatic calibration
system including buffer auto-recognition. The meters include the electrode
and a carry case for travel or storage when not in use. Be sure to ask us
about calibration buffers. Fill out the request form
and ask for more information or call for same day shipping.
More field portable pH meters for special applications:
Multiple parameter data logging meter--
HI 9828 link
| |

|
Direct Soil pH Professional
Meter
|
|
Direct Soil pH Measurement Kit
- Test pH directly
in soil, media and compost
- Dual level LCD
displays pH and temperature simultaneously
- Intuitive menu
prompts simplify operation and calibration
The HI 99121 kit has been designed to address the question
of correct and rapid measurement of pH directly in soil.
Every item in this kit has been made to facilitate
measurements in the field without compromising accuracy.
With HI 99121 you can test the pH of soil directly or after
preparation of a diluted sample. In order to measure the pH
directly, the kit includes a plastic auger to perforate the
ground. The HI 1292D pH electrode has been designed for
direct soil measurement. It has a conic, rugged tip and can
be directly inserted in moist or soft soil. The electrode
incorporates a temperature sensor right near the tip to
enable it to measure and compensate for the temperature
quickly. For higher degrees of accuracy, or for stony ground
where the electrode may be damaged, use the specially made
HI 7051M soil preparation solution. This solution is
included in the package to help you measure pH more
precisely by diluting a sample of the soil in it. The
dual-level LCD shows the pH value from -2 to 16 with 0.01
resolution. It also displays the temperature on the
secondary level at the same time. With a menu-driven custom
microprocessor, the user is helped with clear prompts during
operation and calibration.
HI 99121 is supplied complete with HI 1292D rugged carrying
case pH electrode, ground auger, HI 7051M soil preparation
solution, pH 4 and pH 7 buffer sachets, HI 700663 and HI
700664 electrode cleaning solutions, batteries, rugged
carrying case and instructions.
|
|
|
| |
|
Available Accessories: |
|
HI 1292D |
| |
pH electrode in glass, DIN |
|
HI 721312 |
| |
Rugged carrying case |
|
HI 710007 |
| |
Blue rubber boot |
|
HI 710008 |
| |
Orange rubber boot |
|
HI 7051M |
| |
Soil preparation solution |
|
HI 7004L |
| |
pH 4.01 buffer solution, 500 mL |
|
HI 7007L |
| |
pH 7.01 buffer solution, 500 mL |
|
HI 77400P |
| |
Cal. Kit (pH 4 & 7, 20 mL, 5 pcs each) |
|
HI 721319 |
| |
Ground auger |
|
|
|
Specifications: |
|
|
|
Range
|
|
-2.00 to 16.00
pH, -5.0 to 105.0°C or 23.0 to 221.0°F |
|
Resolution
|
|
0.01 pH, 0.1°C
or 0.1°F |
|
Accuracy (@20°C/68°F)
|
|
± 0.02 pH, ±
0.5°C up to 60°C, ± 1°C up to 105°C or ± 1.0°F up to 140°F,
± 2°F up to 221°F |
|
Typical EMC Deviation
|
|
± 0.02 pH, ±
0.2°C or ± 0.4°F |
|
Calibration
|
|
Automatic 1 or 2
points with 2 sets of memorized buffers (4.01, 7.01, 10.01
or 4.01, 6.86, 9.18) |
|
Temp. Compensation
|
|
Automatic |
|
Electrode
|
|
Glass
pH/temperature electrode, with amplifier, DIN conn. And 1 m
(3.3') cable |
|
Battery Type / Life
|
|
3 x 1.5V
AA/approximately 1500 hours of continuous use |
|
Environment
|
|
0 to 50°C (32 to
122°F); RH 100% |
|
Dimensions
|
|
150 x 80 x 36 mm
(5.9 x 3.1 x 1.4") |
|
Weight
|
|
245 g (8.6 oz.) |
|
|
|
|
|
|
|
|
|

|
|
GLP Meter with
"smart electrode" |
|
Waterproof pH/°C Meter with Smart Electrode
-
"Intelligent" electrode technology
-
GLP
capability with password protection
-
LCD
backlight, graphic symbols and messages
HI 98140 is one of the most innovative portable pH meters ever
offered. This instrument features IP 67 waterproof protection,
GLP (Good Laboratory Practice) capabilities and use the HANNA
instruments®' intelligent electrodes. These pH
electrodes incorporate a chip which memorizes the calibration
data performed with a specific instrument. When the
'intelligent' electrode is attached to the meter again, it is
automatically recognized. This technology allows the operator to
optimize time and efficiency with unsurpassed safety. It avoids
erroneous measurements taken in the event the pH electrode is
substituted after calibration. This series of electrodes also
incorporates a temperature sensor, eliminating the need for an
additional temperature probe.
HI 98140 and HI 98150 are supplied complete with HI 1618D smart
pH electrode with internal temperature sensor, pH 4 and pH 7
buffer sachets, electrode cleaning solution, batteries, 12 Vdc
power adapter, rugged carrying case and instructions. |
|
|
|
|
|
Available
Accessories: |
|
HI
1618D |
| |
Intelligent pH
electrode |
|
HI
76405 |
| |
Electrode
holder |
|
HI
7004L |
| |
pH 4.01 buffer
solution, 500 ml |
|
HI
7007L |
| |
pH 7.01 buffer
solution, 500 ml |
|
HI
7010L |
| |
pH 10.01
buffer solution, 500 ml |
|
HI
70300L |
| |
Storage
solution, 460 ml |
|
HI
77400P |
| |
Cal. Kit (pH 4
& 7, 20 ml, 5 pcs each) |
|
HI
92000 |
| |
Windows®
compatible software |
|
HI
920011/9 |
| |
9-pin cable or
PC |
|
|
|
Specifications: |
|
|
|
Range |
|
-4.00 to 19.99
pH, -20.0 to 120.0°C |
|
Resolution |
|
0.01 pH, 0.1°C
(-10 to 120);1°C below -10°C |
|
Accuracy (@20°C/68°F) |
|
±0.01 pH,
±0.4°C (0 to 70°C); ±1°C (outside) |
|
Typical EMC Deviation |
|
±0.02 pH,
±0.4°C |
|
pH
Calibration |
|
Automatic 1 or
2 points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01,
9.18 or 10.01) |
|
Temp.
Compensation |
|
Automatic from
-20 to 120°C (15 to 248°F) |
|
Electrode |
|
Intelligent
electrode with temp. sensor and 1 m (3.3') DIN cable |
|
Computer Interface |
|
RS 232 |
|
Logging Function |
|
Up to 500
measurements with log-on-demand |
|
Input
Impedance |
|
10¹² Ohm |
|
Battery Type / Life |
|
4 x 1.5V AA;
approximately 300 hours of continuous use (without backlight) or
12VDC adapter (included) |
|
Environment |
|
0 to 50°C (32
to 122°F); RH 100% |
|
Dimensions |
|
196 x 80 x 60
mm (7.7 x 3.1 x 2.4") |
|
Weight |
|
500 g (17.6
oz.) |
|
|
|
|
|
|
|
|
|
|
|
|
|

|
|
GLP Meter with "smart
electrode" |
|
Waterproof
pH/ORP/°C Meter with Smart Electrode
-
"Intelligent"
electrode technology for reliable measurements
-
Password
protection secures GLP data
-
Backlit LCD for
low light conditions
-
mV feature
allows sensor diagnotics
HI 98140 and HI 98150 are two of the most innovative portable pH
meters ever offered. These instruments feature IP 67 waterproof
protection, GLP (Good Laboratory Practice) capabilities and use the
HANNA instruments® intelligent electrodes. These pH electrodes
incorporate a chip which memorizes the calibration data performed with
a specific instrument. When the intelligent electrode is attached to
the meter again, it is automatically recognized. This technology
allows the operator to optimize time and efficiency with unsurpassed
safety. It avoids erroneous measurements taken in the event the pH
electrode is substituted after calibration. This series of electrodes
also incorporates a temperature sensor, eliminating the need for an
additional temperature probe.
HI 98140 and HI 98150 are supplied complete with HI 1618D smart pH
electrode with internal temperature sensor, pH 4 and pH 7 buffer
sachets, electrode cleaning solution, batteries, 12 Vdc power adapter,
rugged carrying case and instructions.
|
|
|
|
|
|
Available
Accessories: |
|
HI 1618D |
| |
Intelligent pH
electrode |
|
HI 3619D |
| |
Intelligent ORP
electrode (glass) |
|
HI 3620 |
| |
Intelligent ORP
electrode (PEI) |
|
HI 76405 |
| |
Electrode holder |
|
HI 7004L |
| |
pH 4.01 buffer
solution, 500 ml |
|
HI 7007L |
| |
pH 7.01 buffer
solution, 500 ml |
|
HI 7010 |
| |
pH 10.01 buffer
solution, 500 ml |
|
HI 70300L |
| |
Storage solution,
460 ml |
|
HI 77400P |
| |
Cal. Kit (pH 4 & 7,
20 ml , 5 pcs each) |
|
HI 7863 |
| |
mV cal. adapter used
w/ HI 8427 |
|
HI 92000 |
| |
Windows® compatible
software |
|
HI 920011/9 |
| |
9-pin cable for PC |
|
|
|
Specifications: |
|
|
|
Range |
|
-4.00 to 19.99 pH,
±600.0 mV; ±2000 mV, -20.0 to 120.0°C |
|
Resolution |
|
0.01 pH, 0.1 mV
(±400 mV); 0.2 mV (±600 mV), 0.1°C (-10 to 120); 1°C below -10°C |
|
Accuracy
(@20°C/68°F) |
|
±0.01 pH, ±0.2mV
(±400 mV); ±0.4 mV (±600 mV), ±0.4°C (0 to 70°C); ±1°C (outside) |
|
Typical EMC
Deviation |
|
±0.02 pH, ±1 mV
(within ±600 mV); ±2 mV (outside), ±0.4°C |
|
pH Calibration |
|
Automatic 1 or 2
points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01, 9.18 or
10.01) |
|
mV Calibration |
|
Automatic 2 points
at 0, 350 mV or 3 points at 0, 350, 1900 mV |
|
Temp. Compensation |
|
Automatic from -20
to 120°C (15 to 248°F) |
|
Electrode |
|
Intelligent
electrode with temp. sensor and 1 m (3.3') DIN cable |
|
Computer Interface |
|
RS 232 |
|
Logging Function |
|
Up to 500
measurements with log-on-demand |
|
Input Impedance |
|
10¹² Ohm |
|
Battery Type / Life |
|
4 x 1.5V AA;
approximately 300 hours of continuous use (without backlight) or 12VDC
adapter |
|
Environment |
|
0 to 50°C (32 to
122°F); RH 100% |
|
Dimensions |
|
196 x 80 x 60 mm
(7.7 x 3.1 x 2.4") |
|
Weight |
|
500 g (17.6 oz.) |
|
|
|
|
|
|
|
|

|
|
Automatic data logging |
|
Printing and
Logging pH/Temp Meter with GLP
-
Intelligent
electrode technology for reliable measurements
-
Powerful meter in
a small package
-
Password
protection secures GLP data
The top-of-the-range HI 98230 and HI 98240 pH meters offer many
advanced features engineered into a highly portable meter, with the
added advantage of having a built-in plain paper printer, logging and
log-on demand. Featuring GLP (Good Laboratory Practice) capabilities,
these instruments incorporate the HANNA instruments® intelligent
electrode technology. All these features make HI 98230 and HI 98240
ideal for on site as well as field laboratories. HI 98230 and HI 98240
are the perfect solution for any application where pH measurements
need to be recorded. In fact, data can be stored or printed directly
on demand or automatically. The stored data (up to 100 measurements
on-demand and up to 1500 automatically) can also be downloaded to a PC
with a standard RS232 serial connection and Hanna’s infrared adapter
and HI 92000 Windows® compatible software (not included), which
permits the transfer of data to other programs.
HI 98230 and HI 98240 are supplied complete with HI 1618D smart pH
electrode with internal temperature sensor, pH 4 and pH 7 buffer
sachets, electrode cleaning solution, 5 paper rolls, batteries, 12 Vdc
power adapter, rugged carrying case and instructions.
|
|
|
|
|
|
Available
Accessories: |
|
HI 1618D |
| |
Intelligent pH
electrode |
|
HI 76405 |
| |
Electrode holder |
|
HI 7004L |
| |
pH 4.01 buffer
solution, 500 ml |
|
HI 7007L |
| |
pH 7.01 buffer
solution, 500 ml |
|
HI 7010L |
| |
pH 10.01 buffer
solution, 500 ml |
|
HI 70300L |
| |
Storage solution,
460 ml |
|
HI 77400P |
| |
Cal. Kit (pH 4 & 7,
20 ml, 5 pcs each) |
|
HI 92000 |
| |
Windows® compatible
software |
|
HI 9200/9 |
| |
9-pin infrared
transmitter for PC |
|
|
|
Specifications: |
|
|
|
Range |
|
-4.00 to 19.99 pH,
-10.0 to 120.0°C (14.0 to 248.0°F) |
|
Resolution |
|
0.01 pH, 0.1°C
(0.1°F) |
|
Accuracy
(@20°C/68°F) |
|
±0.01 pH, ±0.5°C
(±1°F) from 0 to 70 (32°F to 158°F); ±1 (±1.8°F) (outside) |
|
Typical EMC
Deviation |
|
±0.03 pH, ±0.8°C
(±1.4°F) |
|
pH Calibration |
|
Automatic 1 or 2
points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01, 9.18 or
10.01) |
|
Temp. Compensation |
|
Automatic from -10
to 120°C (15 to 248°F) |
|
Electrode |
|
Intelligent
electrode with temperature sensor and 1 m (3.3') DIN cable (included) |
|
Computer Interface |
|
Infrared interface
connected to serial RS 232 port |
|
Printing/Logging
Interval |
|
Selectable from 1,
2, 5, 10, 15, 30, 60, 120 and 180 minutes |
|
Input Impedance |
|
10¹² Ohm |
|
Battery Type / Life |
|
4 x 1.5V
AA/approximately 350 hours of continuous use (without backlight) or
12VDC adapter |
|
Environment |
|
0 to 50°C (32 to
122°F); RH 95% |
|
Dimensions |
|
220 x 82 x 66 mm
(8.6 x 3.2 x 2.6") |
|
Weight |
|
500 g (17.6 oz.) |
|
|
|
|
|

|
|
Automatic data logging |
|
Printing and
Logging pH/ORP/Temperature Meter with GLP
-
Intelligent
electrode technology simplifies use with multiple electrodes
-
Robust portable
tool for multi-parameter analysis
-
Password
protection secures GLP data
The top-of-the-range HI 98230 and HI 98240 pH meters offer many
advanced features engineered into a highly portable meter, with the
added advantage of having a built-in plain paper printer, logging and
log-on demand. Featuring GLP (Good Laboratory Practice) capabilities,
these instruments incorporate the HANNA instruments®� intelligent
electrode technology. All these features make HI 98230 and HI 98240
ideal for on site as well as field laboratories. HI 98230 and HI 98240
are the perfect solution for any application where pH measurements
need to be recorded. In fact, data can be stored or printed directly
on demand or automatically. The stored data (up to 100 measurements
on-demand and up to 1500 automatically) can also be downloaded to a PC
with a standard RS232 serial connection and Hanna’s infrared adapter
and HI 92000 Windows® compatible software (not included), which
permits the transfer of data to other programs.
HI 98230 and HI 98240 are supplied complete with HI 1618D smart pH
electrode with internal temperature sensor, pH 4 and pH 7 buffer
sachets, electrode cleaning solution, 5 paper rolls, batteries, 12 Vdc
power adapter, rugged carrying case and instructions.
|
|
|
|
|
|
Available
Accessories: |
|
HI 3620 |
| |
Intelligent ORP
electrode (PEI) |
|
HI 76405 |
| |
Electrode holder |
|
HI 7004L |
| |
pH 4.01 buffer
solution, 500 ml |
|
HI 7007L |
| |
pH 7.01 buffer
solution, 500 ml |
|
HI 7010L |
| |
pH 10.01 buffer
solution, 500 ml |
|
HI 70300L |
| |
Storage solution,
460 ml |
|
HI 77400P |
| |
Cal. Kit (pH 4 & 7,
20 ml, 5 pcs each) |
|
HI 92000 |
| |
Windows® compatible
software |
|
HI 9200/9 |
| |
9-pin infrared
transmitter for PC |
|
HI 1618D |
| |
Intelligent pH
electrode |
|
HI 3619D |
| |
Intelligent ORP
electrode (glass) |
|
|
|
Specifications: |
|
|
|
Range |
|
-4.00 to 19.99 pH,
±400.0 mV; ±2000 mV, -10.0 to 120.0°C (14.0 to 248.0°F) |
|
Resolution |
|
0.01 pH, 0.1 mV
(±400 mV); 1 mV (outside), 0.1°C (0.1°F) |
|
Accuracy
(@20°C/68°F) |
|
±0.01 pH, ±0.5 mV
(±400 mV); ±2 mV (outside), ±0.5°C (±1°F) from 0 to 70 (32°F to
158°F); ±1 (±1.8°F) (outside) |
|
Typical EMC
Deviation |
|
±0.03 pH, ±1 mV
(within ±400 mV); ±2 mV (outside), ±0.8°C (±1.4°F) |
|
pH Calibration |
|
Automatic 1 or 2
points with 5 memorized standard buffers (pH 4.01, 6.86, 7.01, 9.18 or
10.01) |
|
mV Calibration |
|
Automatic 2 points
at 0, 350 mV or 3 points at 0, 350, 1900 mV |
|
Temp. Compensation |
|
Automatic from -10
to 120°C (15 to 248°F) |
|
Electrode |
|
Intelligent
electrode with temperature sensor and 1 m (3.3') DIN cable |
|
Computer Interface |
|
Infrared interface
connected to serial RS 232 port |
|
Printing/Logging
Interval |
|
Selectable from 1,
2, 5, 10, 15, 30, 60, 120 and 180 minutes |
|
Input Impedance |
|
10¹² Ohm |
|
Battery Type / Life |
|
4 x 1.5V
AA/approximately 350 hours of continuous use (without backlight) or
12VDC adapter |
|
Environment |
|
0 to 50°C (32 to
122°F); RH 95% |
|
Dimensions |
|
220 x 82 x 66 mm
(8.6 x 3.2 x 2.6") |
|
Weight |
|
500 g (17.6 oz.) |
|
|
|
|
|

|
|
Very simple to operate |
|
Waterproof pH Meter
and Digital Thermometer
-
Two buttons for
very simple operation
-
Single probe
measures pH and °C
-
Long battery life
for extended use
The HI 9210N heavy-duty pH meter is designed to provide laboratory
results and accuracy under the harshest of industrial conditions. HI
9210N is supplied with a rugged pH electrode which, in addition to the
pH reference system and sensitive bulb, incorporates a temperature
sensor and a miniaturized amplifier. You need to use only one probe
for both measurements! The electrode signal is also pre-amplified and
strengthened to give you better protection against noise and
electromagnetic interference, a common phenomenon in industrial
applications. Calibration is manual and the trimmers are placed behind
a protective cover. This assures that the unit can be calibrated at
any pH buffer value. The protective cover also keeps the meter
watertight and prevents the calibration from being tampered with. Even
though the HI 9210N’s battery life is an incredible 2000 hours, a low
voltage indicator appears on the screen to alert the user that there
is only 20 hours of battery life left. Afterwards, the BEPS (Battery
Error Prevention System) will stop any false readings by shutting the
meter off.
HI 9210N is
supplied complete with HI 1217D gel-filled combination pH/°C
electrode, and 1 m (3.3') DIN cable, 4 x 1.5V AA batteries and
instruction manual.
|
|
|
|
|
|
Available
Accessories: |
|
HI 1217D |
| |
Combination 4-in-1
pH/°C electrode |
|
HI 721317 |
| |
Field carrying case |
|
HI 76405 |
| |
Electrode holder |
|
HI 7004L |
| |
pH 4.01 buffer
solution, 500 ml |
|
HI 7007L |
| |
pH 7.01 buffer
solution, 500 ml |
|
HI 7010L |
| |
pH 10.01 buffer
solution, 500 ml |
|
HI 70300L |
| |
Storage solution,
460 ml |
|
HI 77400P |
| |
Cal. Kit (pH 4 & 7,
20 ml, 5 pcs each) |
|
|
|
Specifications: |
|
|
|
Range |
|
0.00 to 14.00 pH,
0.0 to 70.0°C |
|
Resolution |
|
0.01 pH, 0.1°C |
|
Accuracy
(@20°C/68°F) |
|
±0.02 pH, ±0.5°C |
|
Typical EMC
Deviation |
|
±0.02 pH, ±0.5°C |
|
pH Calibration |
|
Manual 2 points
through trimmers |
|
Offset Calibration |
|
±1 pH |
|
Slope Calibration |
|
From 85 to 105% of
nominal value |
|
Temp. Compensation |
|
Automatic from 0 to
70°C (32 to 158°F) |
|
Electrode |
|
Combination gel
amplified pH/°C electrode with 1 m (3.3') DIN cable |
|
Input Impedance |
|
10¹² Ohm |
|
Battery Type / Life |
|
4 x 1.5V
AA/approximately 2000 hours of continuous use |
|
Environment |
|
0 to 50°C (32 to
122°F); RH 100% |
|
Dimensions |
|
196 x 80 x 60 mm
(7.7 x 3.1 x 2.4") |
|
Weight |
|
320 g (11.3 oz.) |
|
|
|
pH Meters: HI 9024 - HI 9025 |
| |

|
|
Solid, reliable,
workhorse
|
|
Waterproof Field pH Meters
- Measures pH & Temp
or PH, ORP & Temp (HI 9025)
- Simple to Operate
- Ideal for Field
Measurements
HI 9024 and HI 9025 are microprocessor- based pH meters in
waterproof casings. HI 9024 measures pH and temperature, and
HI 9025 can also measure ORP in the mV range. Both meters
are housed in rugged, waterproof, ABS casings that are built
to last. The Auto-Instruction feature guides the user
step-by-step through the calibration process with graphic
symbols. The automatic calibration procedure is simple and
quick and can be performed at 1 or 2 points selected from 5
memorized buffers: pH 4.01, 6.86, 7.01, 9.18, 10.01. A
stability indicator assures a correct calibration.
HI 9024 is supplied complete with HI 1230B pH electrode, HI
7669/2W temperature probe, pH 4 and pH 7 buffer sachets, 100
mL plastic beaker, batteries, rugged carrying case and
instructions. HI 9025 is supplied complete with HI 1230B pH
electrode, HI 7669/2W temperature probe, pH 4 and pH 7
buffer sachets, 100 mL plastic beaker, batteries, rugged
carrying case and instructions.
|
|
|
| |
|
Available Accessories: |
|
HI 1230B |
| |
pH elec., gel filled, PEI, BNC |
|
HI 3230B |
| |
ORP elec., gel filled, PEI®, BNC |
|
HI 7669/2W |
| |
Temperature probe |
|
HI 76405 |
| |
Electrode holder |
|
HI 7004L |
| |
pH 4.01 buffer solution, 500 mL |
|
HI 7006L |
| |
pH 6.86 buffer solution, 500 mL |
|
HI 7007L |
| |
pH 7.01 buffer solution, 500 mL |
|
HI 7009L |
| |
pH 9.18 buffer solution, 500 mL |
|
HI 7010L |
| |
pH 10.01 buffer solution, 500 mL |
|
HI 70300L |
| |
Electrode storage solution, 500 mL |
|
HI 8427 |
| |
pH/mV electrode simulator |
|
HI 931001 |
| |
pH/mV electrode simulator with display |
|
|
|
Specifications: |
|
|
|
Range pH
|
|
0.00 to 14.00 pH |
|
Range mV (HI 9025 Only)
|
|
±399.9 mV; ±1999
mV |
|
Range Temp
|
|
0.0 to 100.0°C /
32.0°F to 212.0°F |
|
Resolution pH
|
|
0.01 pH |
|
Resolution mV (HI 9025 Only)
|
|
0.1 mV; 1 mV |
|
Resolution Temp
|
|
0.1°C / 0.1°F |
|
Accuracy (@20°C/68°F) pH
|
|
±0.01 pH |
|
Accuracy (@20°C/68°F) mV (HI
9025 Only)
|
|
±0.2 mV; ±1 mV |
|
Accuracy (@20°C/68°F) Temp
|
|
±0.4°C / ±0.8°F |
|
pH Calibration
|
|
automatic, 1 or
2 point with 5 memorized buffer values (pH 4.01, 6.86, 7.01,
9.18, 10.01); offset calibration: ±1 pH; slope calibration:
70 to 108% |
|
Temp. Compensation
|
|
automatic or
manual, -5 to 105°C (23°F to 221°F) |
|
pH Electrode
|
|
HI 1230B, double
junction, PEI® body, gel filled, BNC connector and 1 m cable
(included) |
|
Temp. Probe
|
|
HI 7669/2W with
1 m (3.3ft) cable (included) |
|
Input Impedance
|
|
1012 Ohm |
|
Battery Type / Life
|
|
4 x 1.5V AA /
approx. 500 hours of continuous use; auto-off after 20
minutes of non-use (can be disabled) |
|
Environment
|
|
0 to 50°C (32 to
122°F); RH max 100% |
|
Dimensions
|
|
196 x 80 x 60 mm
(7.7 x 3.1 x 2.4inches) |
|
Weight
|
|
500 g (1.1 lb.) |
|
|
|
|
|
| |

|
Measures pH & ORP with
one probe
$350
|
|
Portable pH Meter with Sensor Check
- Ideal for Water
Plating baths, potable water & Swimming Pools
- 4-in-1 Probe (pH,
ORP, pH/mV & temperature)
- Sensor Check
Feature
HI 991003 is the first pH meter that can measure pH, pH-mV
(to check the electrode status), ORP and temperature with
only one probe. Since ORP measurements are strongly affected
by pH level, HANNA instruments ® has developed the first
meter that simultaneously measures both ranges. With all
other instruments you need to switch the pH electrode with
an ORP electrode (or vice-versa) and, therefore, it is
impossible to have both measurements at the same time. HI
991003 is largely used to control plating bath waste water,
swimming pool water quality and potable water sanitizing
processes. HI 991003 is provided with our unique Sensor
Check feature that allows the user to establish the status
of the pH electrode before performing measurements. The
Sensor Check feature consists of a pH-mV scale that shows
the characteristics (slope and offset) of the sensor,
allowing the user to determine if the electrode is in good
condition. HI 991003 is protected in a waterproof casing for
use in humid and wet environments. It also features a large
duallevel LCD that displays pH or ORP and temperature
readings simultaneously.
HI 991003 is supplied complete with HI 1297D pH/ORP
electrode with internal temperature sensor, pH 4 and pH 7
buffer sachets, HI 700661 electrode cleaning solution (2 x
20 mL), batteries, rugged carrying case and instructions.
|
|
|
| |
|
Available Accessories: |
|
HI 1297D |
| |
Combo amplified pH/ORP/Temp electrode |
|
HI 721312 |
| |
Hard carrying case |
|
HI 710007 |
| |
Blue shockproof rubber boot |
|
HI 7004L |
| |
pH 4.01 buffer solution, 500 mL |
|
HI 7007L |
| |
pH 7.01 buffer solution, 500 mL |
|
HI 7010L |
| |
pH 10.01 buffer solution, 500 mL |
|
HI 7020L |
| |
Test solution, 200-275 mV, 500 mL |
|
|
|
Specifications: |
|
|
|
Range
|
|
-2.00 to 16.00
pH, ±825 pH (in mV), ±1999 mV, -5.0 to 105.0°C or 23.0 to
221.0°F |
|
Resolution
|
|
0.01 pH, 1 pH
(in mV), 1 mV, 0.1°C or 0.1°F |
|
Accuracy (@20°C/68°F)
|
|
±0.02 pH, ±1 pH
(in mV), ±2 mV, ±0.5 up to 60°C, ±1 up to 105°C, ±1 up to
140°F, ±2 up to 221°F |
|
Typical EMC Deviation
|
|
±0.02 pH, ±2 pH
(in mV), ±2 mV, ±0.2°C or ±0.4°F |
|
pH Calibration
|
|
Automatic 1 or 2
points with 2 sets of standard buffers (4.01, 7.01, 10.01 or
4.01, 6.86, 9.18) |
|
Temp. Compensation
|
|
Automatic from
-5 to 105°C (23 to 221°F) (for pH only) |
|
Electrode
|
|
Combination
amplified pH/ORP/temperature with DIN connector and 1 m
(3.3') cable |
|
Battery Type / Life
|
|
3 x 1.5V
AA/approximately 1500 hours of continuous use; auto-off
after 8 minutes |
|
Environment
|
|
0 to 50°C (32 to
122°F); RH 100% |
|
Dimensions
|
|
150 x 80 x 36 mm
(5.9 x 3.1 x 1.4") |
|
Weight
|
|
245 g (8.6 oz.) |
|
|
|
|
|
| |

|
Flat tip probe for
surface measurements
|
|
Leather & Paper pH Meter
- Important for
Production & Packaging
- Quick & Easy
Measurements
- Flat Tip Probe
Designed for Surface Contact
Direct measurement of pH on paper and leather is made fast
and easy with the HI 99171 pH meter provided with a flat tip
probe, specially designed to optimize surface contact with
the sample. pH measurement of papers and cartons is
important, not only in the production phase, but also in the
packaging phase. The food industry, for example, performs
many compatibility tests; one of these is pH. Measurements
are automatically compensated for the effects of temperature
with the internal sensor of the probe and the LCD
continuously displays the sample's temperature. All
operations can be performed simply with 2 buttons; messages
and symbols on the display guide the user through each
measurement and calibration procedures. pH and temperature
ranges are extended, from -2.00 to 16.00 pH and -5.0 to
105.0 °C respectively. Calibration is automatic at 1 or 2
points, selectable by the user between 2 memorized buffer
sets. Battery life is more than 1500 hours and the
percentage of the remaining charge is displayed when the
meter is turned on. HI 99171 turns off automatically after 8
minutes of inactivity.
HI 99171 is supplied with HI 1414D pH electrode, pH 4 and pH
7 buffer sachets, HI 700680 cleaning solution (2 x 20 mL
sachet), HI 70960 sample preparation solution (30 mL),
batteries, rugged carrying case and instructions.
|
|
|
| |
|
Available Accessories: |
|
HI 1414D |
| |
Amplified flat tip pH electrode |
|
HI 7004L |
| |
pH 4.01 buffer solution, 500 mL |
|
HI 7007L |
| |
pH 7.01 buffer solution, 500 mL |
|
HI 7010L |
| |
pH 10.01 buffer solution, 500 mL |
|
HI 70004P |
| |
pH 4.01 buffer sol., 20 mL |
|
HI 70007P |
| |
pH 7.01 buffer sol., 20 mL |
|
HI 70010P |
| |
pH 10.01 buffer sol., 20 mL |
|
HI 77400P |
| |
pH 4 & pH 7 buffer sachets, 20 mL |
|
HI 70300L |
| |
Electrode storage solution, 500 mL |
|
HI 70960 |
| |
Electrolyte sol. for sample prep, 30
mL |
|
HI 700680P |
| |
Cleaning sol. for cellulose dep. 20 mL |
|
HI 710007 |
| |
Shockproof rubber boot, blue |
|
HI 710008 |
| |
Shockproof rubber boot, orange |
|
|
|
Specifications: |
|
|
|
Range
|
|
-2.00 to 16.00
pH |
|
Range
|
|
-5.0 to 105.0°C
/ 23.0 to 221.0°F |
|
Resolution
|
|
0.01 pH |
|
Resolution
|
|
0.1°C / 0.1°F |
|
Accuracy (@20°C/68°F)
|
|
±0.02 pH |
|
Accuracy (@20°C/68°F)
|
|
±0.5°C (up to
60°C), ±1°C (outside); ±1.0°F (up to 140°F), ±2°F (outside) |
|
pH Calibration
|
|
automatic, 1 or
2 point with 2 sets of memorized buffers (pH 4.01/7.01/10.01
or 4.01/6.86/9.18) |
|
Temp. Compensation
|
|
automatic, -5 to
105°C (23 to 221°F) |
|
pH Electrode
|
|
HI 1414D,
amplified, flat tip, with internal temperature sensor
(included) |
|
Battery Type / Life
|
|
3 x 1.5V AA /
approx. 1500 hours of continuous use; auto-off after 8
minutes of non-use |
|
Environment
|
|
0 to 50°C (32 to
122°F); RH max 100% |
|
Dimensions
|
|
150 x 80 x 36 mm |
|
Weight
|
|
210 g (7.4 oz.) |
|
|
|
|
|
|